Instructions for operating the metallographic specimen embedding machine
来源: | 作者:admin | Date :2024-06-10 | 2488 views | 🔊 Click to read aloud ❚❚ ▶ | Share:

Metallographic specimen embedding machine is a pre process that involves embedding and shaping small or irregular shaped and difficult to handle specimens using thermosetting plastics, and is also an essential step in grinding and polishing work.


The metallographic embedding machine has the advantages of low heat release during solidification, low thermal shrinkage correction, and good heat resistance. It is suitable for the use of micro slice solidification adhesive in the electronics industry and is suitable for cold embedding of various small metallographic specimens. Suitable for pressing thermosetting plastics on small metallographic specimens that are not shaped or easy to handle. After forming, it is convenient to perform sample grinding and polishing operations, and it is also beneficial for microstructure determination under a metallographic microscope.


Instructions for operating steps:


1. Turn on the power switch, and then carry out the relevant preparation work. It is necessary to check whether there is any electric wood powder sticking to the edges of the upper and lower modules, and immediately remove it (it can be removed by a woodworking saw blade without excessive force to prevent scratching the control module).


2. If the temperature rises to the set temperature (usually 130-140 ℃), the actual operation can be gradually carried out afterwards.


3. Adjust the handwheel to make the lower mold parallel to the lower service platform; Subsequently, place the specimen facing downwards at the lower mold management center and rotate the handwheel 10-12 times in the opposite direction to move the lower mold and specimen downwards (the height to width ratio of the specimen should generally not exceed one centimeter). Add filling material to make it parallel to the lower service platform, then press the upper mold onto the filling material, apply a downward force on the upper mold with your left finger, and rotate the handwheel in the opposite direction with your left hand to lower the lower mold to a surface smaller than the upper service platform.


4. Quickly close the cover plate and turn the octagonal knob clockwise to make the screw on the knob press against the lower mold, with a slightly strong touch. Subsequently, immediately rotate the handwheel clockwise until the work pressure light is on. At this point, add 1-2 more turns. Do not add too much, as excessive work pressure can cause difficulty in sampling and result in explosion.


5. After the insulation indicator light flashes, under all normal working pressure requirements, hold for five minutes, and the test sample embedding is successful.


6. The entire sampling process is as follows: first, rotate the handwheel in the opposite direction to release pressure until the working pressure light goes out. Then, rotate it another 3-5 turns. Next, rotate the octagonal knob clockwise to push down the upper control module and remove the sample from the mold. Finally, rotate the octagonal knob in the opposite direction to open the cover plate.


7. Rotate the handwheel clockwise to shoot the lower mold. When the top is higher than the lower edge of all molds and parallel to the lower service platform, use a long block to search for the lower mold and place it in the corner of the upper service platform on the right side of the inlaying machine cavity.


8. Once again, slowly raise the lower mold and expose all the test samples before removing them, using the same method as removing the lower mold.


9. Before embedding each sample, it is necessary to pay attention to removing the edges of the upper and lower modules to ensure smooth embedding.